Product Specifications
-
User defined “logging” intervals as frequently as 10 seconds
-
AC (90 – 240 V) and DC Power
-
Rechargeable battery – Approximately 12 hours of “runtime”
-
Password protected menu options
-
Easy calibration
-
Web-page enabled real-time view
-
Ethernet and USB serial communications
-
16-Bit A/D converter
-
1 gigabyte of on-board storage
-
SDS Dimensions: 16”(40.64cm)D x 20”(50.8cm)L x 8”(20.32cm)H
-
SDS Weight: 29 lbs (13.15kg)
-
Shipping Carton Dimensions: 25”(63.5cm)D x 22”(55.88cm)L x 15”(38.1)H
-
Shipping Carton w/SDS Weight: 35 lbs (15.87kg)
|
Degrees F. |
A/D |
Accuracy |
Accuracy |
Accuracy |
|
TC Type |
Min |
Max |
Full scale |
Full Scale |
At 1600 Degrees F. |
At 1600 Degrees F with thermistor error |
B |
32 |
3308 |
0.0124% |
0.0179% |
0.29 |
0.65 |
C |
32 |
4208 |
0.0062% |
0.0067% |
0.11 |
0.47 |
E |
-328 |
1832 |
0.0031% |
0.0032% |
0.05 |
0.41 |
J |
-346 |
2192 |
0.0031% |
0.0036% |
0.06 |
0.42 |
K |
-328 |
2502 |
0.0031% |
0.0045% |
0.07 |
0.43 |
N |
-328 |
2372 |
0.0031% |
0.0052% |
0.08 |
0.44 |
NNM |
0 |
1409 |
0.0031% |
0.0033% |
0.05 |
0.41 |
R |
-58 |
3214 |
0.0062% |
0.0118% |
0.19 |
0.66 |
S |
-58 |
3214 |
0.0124% |
0.0133% |
0.21 |
0.57 |
T |
-328 |
752 |
0.0062% |
0.0119% |
0.19 |
0.55 |
Table 1 - Thermocouple Types, Scaling, and Accuracy